LASERGASTM II Multipass
Monitor
The LaserGas
II Multipass monitor is a high performance gas monitor specially designed for
measurement of very low gas concentrations. Based on the well-proven tuneable diode
laser (TDLAS) technology the instrument has a number of unique features.
·
Short
response time (typically 5-10 sec)
·
No
cross interference from other gases
·
Very
low detection limits (ppb for most gases)
·
Offline
gas analysis in controlled environment
·
Turn-key
instrument
·
Low
cost of ownership
·
Ethernet
connection
The consists of
5 modules, the transmitter, receiver, Multipass cell, beam alignment chamber
and power supply unit, which are all integrated on a wall mounted panel for
easy installation. The standard Multipass cell is made of stainless steel.
Special (e.g. heatable) versions of these cells are available on request.
Measurement readings are provided through standard 4-20 mA current loops, a
fibre-optic output or the optional Ethernet interface. The latter enables
communication via a local area network or the Internet.
Measuring
principle
Unlike
conventional UV or IR spectrographic instruments, LaserGas II Monitors employ
the measurement principle known as ‘single line spectroscopy’,
which eliminates cross interference from other gases. A single gas absorption
line with no cross interference is chosen in the near IR spectral range and
scanned by a single-mode diode laser. The laser beam is coupled into the
measurement cell, a so-called Herriot cell, where it is reflected multiple
times between two spherical mirrors. Subsequently a detector collects the
transmitted light for further analysis and calculation of the gas
concentration. With the Multipass concept a long optical path length (and low
detection limit) can be achieved with a very compact instrument.
Installation
and operation
The
LaserGas II MP Monitor is a turn-key instrument, delivered with factory
calibration and tuned to the specific application. No other operations than
connecting sample gas tubes and optional purge are required during
installation. To avoid fouling of optical parts in the Multipass cell the
cleanliness of the sample gas must be ensured.
Filtering the sample gas in an appropriate extractive system may be required
for some applications. The instrument contains no moving parts, thereby
limiting preventive maintenance to visual inspection. Calibration checks are
recommended every 6-12 months.
Main
applications
The
LaserGas II MP is designed for the detection of very low gases concentrations
(down to ppb level).
The following are some of the typical applications:
·
Trace
measurement of contaminants (e.g. O2, H2O) in pure gases
·
Trace
measurements of contaminants in reactive gases
used in the semiconductor industry
·
Measurement
of contaminants in natural gas
·
Low
ppm H2S measurement in process gases
·
ppm
NO measurement for emission control
Table of Gases
|
Gas
|
Detection
limit
|
Min. meas.
range
|
|
O2
|
1 mg/m3
|
0 - 100 mg/m3
|
|
H2O
|
0,003 mg/m3
|
0 - 0,3 mg/m3 *
|
|
H2S
|
0,3 mg/m3
|
0 - 30 mg/m3
|
|
CH4
|
0,02 mg/m3
|
0 - 2 mg/m3
|
|
CO
|
3 mg/m3
|
0 - 300 mg/m3
|
|
CO2
|
3 mg/m3
|
0 - 300 mg/m3
|
|
NO
|
1,5 mg/m3
|
0 - 150 mg/m3
|
|
N2O
|
0,5 mg/m3
|
0 - 50 mg/m3
|
|
NH3
|
0,015 mg/m3
|
0 - 1,5 mg/m3
|
|
HCl
|
0,005 mg/m3
|
0 - 0,5 mg/m3
|
|
HF
|
0,0015 mg/m3
|
0 - 0,15 mg/m3
|
|
HCN
|
0,03 mg/m3
|
0 - 3 mg/m3
|
Remark: Detection limits are specified for gas temperature
/ pressure = 25°C / 1 bar abs. (* lowest detection limit for H2O
may only be achievable with special sealing of the cell). Other gases and
ultra-low CO and CO2 (range 0-3 mg/m3) are available on